Clarity INT9 A/D converter Code/Rev.: M105/25B – 11 July 2007 Phone: +420 - 251 013 400 © DataApex Ltd. 2007 Fax: +420 - 25
Clarity - INT9 A/D converter Installation 10 6. Enter the detector names in Name fields for individual channels, set signal units. Note: A detaile
Clarity - INT9 A/D converter Installation 11 2.4.1 Measuring on multiple Instruments Rather than dragging the entire INT9 A/D Card Driver icon from
Clarity - INT9 A/D converter Installation 12 2.4.2 Using multiple INT9 cards When using multiple A/D cards, it will be necessary to add the INT9 ca
Clarity - INT9 A/D converter Installation 13 Device Manager, otherwise you will not be able to add it in the System Configuration dialog and configu
Clarity - INT9 A/D converter Using the INT9 card 14 3 Using the INT9 card There are generally two places for setting the parameters of INT9 A/D Co
Clarity - INT9 A/D converter Using the INT9 card 15 Channel 1 (to 4) For each channel of the INT9 A/D Converter the name of the signal can be edite
Clarity - INT9 A/D converter Using the INT9 card 16 Supply Frequency Allows to change the Mains Supply Frequency Suppression. It is used for suppre
Clarity - INT9 A/D converter Using the INT9 card 17 Select Detector Selects the detector whose parameters will be set. If there is only one detecto
Clarity - INT9 A/D converter Using the INT9 card 18 3.2.2 Method Setup - Measurement Fig. 11. Method Setup - Measurement External Start/Stop Ena
Clarity - INT9 A/D converter Using the INT9 card 19 Down The station reacts to changes in voltage in the relevant controlling input from high (>
Sections of the manual connected only to the Clarity Full version are marked with the icon and grey stripe to the right of the text.
Clarity - INT9 A/D converter Troubleshooting 20 4 Troubleshooting If you will not find your answers here, use the www.dataapex.com website where the
Clarity - INT9 A/D converter Troubleshooting 21 4.2 Problems with INT9 driver • “Board malfunction” If the error message "Board malfunction&qu
Clarity - INT9 A/D converter Troubleshooting 22 4.2.1 How to check the INT9 driver: How to check the INT9 driver: Windows 2000/XP/ Vista In the St
Clarity - INT9 A/D converter Troubleshooting 23 • The item does appear and there are no error symbols around it. A driver for a different type of W
Clarity - INT9 A/D converter Troubleshooting 24 • Select the Install from a list or specific location (Advanced) option. • Select Include this lo
Clarity - INT9 A/D converter Troubleshooting 25 • In the System Restore dialog use the Restore my computer to an earlier time option and press the N
Clarity - INT9 A/D converter Troubleshooting 26 Caution! This procedure can only be used safely if no other hardware has been installed since the i
Clarity - INT9 A/D converter Troubleshooting 27 • Select Add/Troubleshoot a device and press the Next button. • In the Devices list select Add a
Clarity - INT9 A/D converter Troubleshooting 28 • Select No, I want to select the hardware from a list and press the Next button. • In the Hardwa
Clarity - INT9 A/D converter Troubleshooting 29 • In the Models list select the DataApex Acquisition device, Model 9.00 and press the Next button.
Clarity - INT9 A/D converter Table of Contents 3 1 Brief Description...4 1.1 INT9 Acquis
Clarity - INT9 A/D converter Troubleshooting 30 If your A/D card is not in the left-hand Setup Control Modules list, add it using the Add button and
Clarity - INT9 A/D converter Troubleshooting 31 If you only have Demo detectors in the left-hand Setup Control Modules list and your A/D card is mi
Clarity - INT9 A/D converter Tables and specifications 32 5 Tables and specifications 5.1 Description of the INT9 card connector (Male) Connector
Clarity - INT9 A/D converter Tables and specifications 33 5.2 Parameters of digital Inputs and Outputs The card contains eight digital TTL outputs
Clarity - INT9 A/D converter Tables and specifications 34 5.3 INT9 - CE Conformity Declaration
Clarity - INT9 A/D converter Tables and specifications 35 5.4 Technical Data Converter type: 24-bit ΔΣ (delta – sigma) Analog inputs: differentia
Clarity - INT9 A/D converter Brief Description 4 1 Brief Description This manual describes the use of the INT9 converter with Clarity software ver.
Clarity - INT9 A/D converter Brief Description 5 A/D Converter types in general Prior to collecting and processing the chroma-tographic signal by a
Clarity - INT9 A/D converter Installation 6 2 Installation Ensure that you have Administrator access rights to your Windows OS before you proceed wi
Clarity - INT9 A/D converter Installation 7 and tinned endings – red (+), white (-) and shielding (analogue ground)]. • Starting (marker) cables Lab
Clarity - INT9 A/D converter Installation 8 Starting cableSignal cableSTART (REMOTE, …)RedShieldingSYMMETRICAL CONNECTION+ (+1V, Hi, …)- (-1V
Clarity - INT9 A/D converter Installation 9 2.3.2 Connection of starting cables: Starting input reacts to a change of the TTL logical level (5V) or
Commenti su questo manuale